Test Type Packages

Choose From a Standard ProCalV5 TTP or Have One Customized

ProCalV5 Test Type Packages are designed to support advanced paperless functionality with comprehensive testing and maintenance routines for a full range of complex systems and devices. All Test Type Packages shown below are already built and can be supplied as add-ons to your existing ProCalV5 installation.

Our engineers can also customize the existing packages below to your particular company needs or build others to match your unique requirements.

All Test Type Packages are validated, 21 CFR part 11 compliant and developed within our strict, audited practices.

Air Particle Counters Records all data from offsite vendor calibration

Air Sampler 1 Test aspirated volume & record K3 factor with a MAS-100C calibrator

Air Sampler 2 Test aspirated volume & record K3 & K5 factor with a DA-100 calibrator

Air Sampler 3 Test differential pressure; look up related flow value in a table

Bio Containment Cabinet This test type performs a test for Downflow Velocity, Inflow Velocity, HEPA Filter Integrity, as well as, performance tests for Cabinet Integrity, Air Flow Smoke Pattern, Site Installation Assessment, and Electrical Tests.

Bio Safety Cabinet This TTP allows users to certify Bio Safety Cabinets.  On the visual prompt provided in ProCalV5, users can record Downflow Velocity, Test Method, HEPA Filter Integrity, Air Flow Smoke and an Acceptance Criteria.  Pass/fail is determined by HEPA Integrity Result, Smoke Test Downflow Uniformity, Smoke Test Containment, Smoke Test Opening Containment, Downflow Velocity Average (Standard Test), Back, Middle and Front Velocity Averages (Zoned Test) and Intake Velocity Average.

Calibration Codes The Calibration Codes performs normal ProCal calibration calculations and then assigns a letter A, B, C, or D as the output based on the results

Chart Recorders Test the accuracy of the timing functionality of the chart recorder

Chemistry ScreenChemistry The Dissolution – Chemistry provides for Tablet Calibration test portion of Dissolution Bath calibrations.  This TTP results in the eventual comparison of a “% Dissolved” value for each vessel, compared to the minimum and maximum % Dissolved based on a USP Lot number.

Divisions Verification – Force Calibration The Divisions Verification interpolates intermediary values of the output signal type between test readings, based on the difference of those test readings divided by the difference in the input set points. No Pass/Fail is determined by this test but it can be used for reporting purposes

Extreme Values Test The Extreme Values accepts a set of test point inputs and calculates the minimum or maximum reading from the set

Fume Hood Certification This TTP allows users to certify fume hoods.  User set the Output Resolution and the Acceptable Low & High Values.  Also a user can set the # of Columns, # of Rows, Sash Opening and Sash Width. During calibration is where the Average Velocity, Hood Opening, and Air Volume are calculated and the Pass/Fail is based upon the average velocity compared to a user established acceptable specified range.

Gage R&R Study The ProCalV5 Gage R&R Utility allows users to collect data for a Gage Repeatability and Reproducibility study on individual items they are managing and report on range of operators, acceptability and more.

Glove Box CertificationGlove Box Certification This will allow users to perform glove box certifications.  Box Height and Sash Width for the glove box are entered on the item’s Master Instrument Record.  The visual prompt on the Calibration screen is when a user performs a calibration/certification to their Glove Box.

HEPA Filter Integrity The HEPA Filter test will summarize the results for HEPA Filter Integrity tests (in VLF mode) and apply a pass/fail to the data.  The test will calculate the Average Face Velocity = Mean of all the entered Velocity Data entered in any preceding test point groups that use the “VLF” “Test Type” input value and determine pass/fail based on the Low Velocity & High Velocity calibration points.

HPLC Autosampler The HPLC Autosampler involves the gathering of two vial weight readings from a HPLC Autosampler to be compared to some expected weight change within a tolerance specified as Pct of Reading

HPLC Extreme Values Test ScreenHPLC Extreme Values Test The Extreme Values accepts a set of test point inputs and calculates the minimum or maximum reading from the set from the results of the HPLC Detector

HPLC Pump Flow Rate  The HPLC Pump Flow Rate provides capability for using a Flow Meter, a Class A Buret, or a Volumetric Flask to calibrate the Flow Rate of a HPLC Pump

HPLC Pump Gradient Proportioning The HPLC Pump Gradient Proportioning provides capability for determining the percent flow contribution from several channels of a HPLC Pump

Input Conversion This will allow users to display the As Found and As Left readings back in the unit of the Input Signal Type in ProCalV5.  This test defaults to a linear correlation from Input Range to Output Range unless the Correlation to Output is specifically set to Square Root by a user.

Laminar HEPA Filter Test The Laminar HEPA Filter Test calculates the Average Flow and a limit based on that Average Flow across a HEPA Filter from a set of flow measurements

Linearity Test ScreenLinearity Test The Linearity calculates a correlation coefficient and its square from a set of inputs and readings and compares them against a standard of acceptance

Manual Pass/Fail The Manual Pass/Fail allows the user to identify the test using the Group Name field, and enter whether or not it passed or failed as well as being able to record an As Found OOT

Metric Weights Test (Standard Deviation) Test average differential, calculate average weight & Mean Standard Deviation

Monitor Mark/Freeze Watches Two tests at two different temperatures for a time period

Multiple Readings The multiple readings test will allow users to enter multiple As Found and As Left readings for a single input reading.  The user can predefine a set of Heading Labels and Units for up to 5 additional As Found and As Left columns related to their test points.  All values are captured on the Calibration record in ProCalV5.

PE 900 Series Interface The PE 900 Series Interface captures three values for each of two Channels (A and B): Percent Gain, SD Gain Error and Linearity Error according to a +/- deviation from zero Tolerance specification

Pipette Calibration This allows users to capture multiple repeated volumes and weights per each Pipette test volume. Each set of readings is tested for precision (%RSD) and Accuracy (% of target volume).

Pipette Calibration ScreenPipette Calibration with Conversion This test type allows users to capture multiple repeated weights per each Pipette test volume, and convert to the corresponding volume according to an entered temperature value. This is done using a water density factor table. Conversion from milliliter and microliter to milligrams or grams is done automatically. Each set of readings is tested for precision (%RSD) and Accuracy (% of target volume).

Pipette Calibration Z Factor This test type performs a conversion between weight and volume. A water density table is utilized for the conversion calculations. The water density table is shown in Attachment 1. This Pipette Test Type tests the accuracy at each reading and the Mean Volume, Accuracy Percentage, and Precision Percentage for each set of test volumes.

Pressure Safety Valves This Test Type allows for the calibration of your Pressure Safety Valves including a thorough checklist of As Found condition, Critical Dimensions, Repairs and Final Results.

Reproducibility ScreenReproducibility The Reproducibility TTP calculates the Standard Deviation and Reproducibility value for the given test group, for a variable number of readings, a standard value, and a maximum Reproducibility value.  It then compares the Reproducibility to the maximum Reproducibility value to determine the Pass/Fail.

RTD Curve Temperature-to-Ohms conversion in Celsius and Fahrenheit, lookup or interpolate Inputs from Outputs or Outputs from Inputs, from Temperature to Ohms and Ohms to Temperature {DIN Platinum RTD Temp/Resistance Tables, Curve PR-238 (°C vs Absolute Ω) and Curve PR-239 (°F vs Absolute Ω)}.

Scales & Balances ScreenScales & Balances Multiple tests; Zero test, Corner Test & Repeatability

SMA Sampler This will perform Pass/Fail on a Velocity Reading and a Time reading, in fractional Minutes.  It will also give a Replace Battery warning when the Average Voltage of a set of test point readings is outside of a given acceptable range.

Stopwatch Test The Stopwatch Test calculates the Deviation, % Deviation and % Error for a Stopwatch based on a specified standard frequency and a reading of a measured frequency.

Stopwatch Timer Test The Stopwatch Timer Test allows the user to capture and automatically process and calculate time measurements to be compared with some target time. Time can be measured in Days, Hours, Minutes, Seconds and up to 100 millionth of a second (8 decimal places)

Switch Test ScreenSwitch Test The Switch Test allows the user to capture either a single point trip test or a two-point trip and reset test. Many of the features of ProCalV5’s standard calibration are used, such as the variety of limit calculations and providing the ramp range. A special feature of this Test Type package is the ability to download this test to a Field Calibrator using the ProCalV5 Field Calibrator Utility FCINTF interface.

Tablet Press The Tablet Press calculates a Tab / Hr * 1000 value for a number Tablet Press stations and compares it to the indicated Tab / Hr * 1000 value

Task List This will supply and record a list of tasks for maintenance or calibration-related maintenance steps to be recorded on a Calibration record. Each test point will be a “Task”. During Calibration the user can mark whether a task has been completed successfully or not, along with recording a brief comment for each one. The group can be set up to fail the calibration if any step is skipped, and/or if any step’s result is not filled in, but may also be set up to be for information only.

Temperature The Dissolution – Temperature allows the user to capture a given number of temperature test points, each one representing a vessel in a Dissolution Bath, or an External Bath, or some Other Probe related to the test

Template Group Load Template Group Load Test Type Package will allow users to load a test point group from a different item dynamically during calibration entry.  This allows certain items to be used as “templates” for groups in another item.  Optionally, only a certain number of points will be taken from the requested item/test point group as well.

Test Omission The Test Omission allows the user to omit a particular test group from the overall determination of Calibration Pass/Fail

Timer Test The Timer Test allows the user to capture and automatically process and calculate time measurements to be compared with some target time recording in either Minutes or Seconds

Uncertainty Calculation ScreenUncertainty Calculation Module
Calculate for Type A or B
Document Uncertainty Contributors for Instrument Under Test
Document Uncertainty Contributor for Test Standard Used
Calculate Combined Uncertainty

Wobble The Dissolution – Wobble allows the user to capture the Apparatus used when performing Dissolution Wobble tests of several vessels of a Dissolution Bath

Learn More About ProCalV5
Click Here to View the ProCal Brochure
Download the ProCal Features eBook by Clicking Here
Request a Free ProCal 30 Day TrialGet a Free Quote on ProCal Calibration Software
<alt="Click Above To Learn More About ProCalV5">

Click here for more information on ProCal Calibration Software